Tuesday, 11 September 2018

"The quality of the results obtained with this product are sub-optimal. Not so great!"

PRODUCT:                 DRAQ5

Application Area:       Analyzed nuclei in Drosophila ovaries using Draq5


Ease of Use:              ⭐⭐⭐⭐⭐
After Sales Service:    ⭐⭐
Value for Money:        ⭐⭐⭐
 Average rating:          4.0


Comments: The product itself is easy to use but the results are not reproducible every time.  Would love if the product quality is improved because overall it is a good reagent to have!

Name: 
Tanu Singh

Organisation: Fox Chase Cancer Center

.........................................

Manufacturer's response: The initial review posting elicited an immediate technical support response. After successful testing of both fresh DRAQ5 and DRAQfx (for comparison) in the customer’s lab the issue was isolated to the presence of 0.3% Triton X-100 during the final counterstaining step. This was remedied by an intermediate wash with PBS. This learning will be reflected in our products’ technical data sheets for fixed cell immunofluorescence protocols.